Continuous screening of health record data in adults (18+) from emergency arrival or admission through the acute-care stay. It shows a ‘Sepsis Risk High’ flag in the record to help clinicians detect sepsis within 24 hours; it must not be the sole basis for diagnosis.
Not directly applicable — a clinical alerting system, not a regulated diagnostic device in the FDA/CE sense.
Adams et al., Nature Medicine 2022 (founder co-author): 5 US hospitals, 6,877 sepsis patients; alert confirmed within 3 h linked to 3.3-point lower adjusted mortality, not randomised.
FDA 510(k) data: retrospective, 7,732 encounters, 3 US sites; sensitivity 79.4%, specificity 89.5%, 11.7% of flags were true sepsis.
Schinkel, van der Poll, Wiersinga (Amsterdam UMC), Am J Respir Crit Care Med 2023, commentary: observational design and record-based sepsis labels make the benefit hard to judge; results seem preliminary.
No independent study of TREWS accuracy or patient outcomes outside the developers’ team was found.
Deployed at five Johns Hopkins hospitals, with among the strongest quantified clinical-impact figures found in this scan.
– Mortality benefit comes from observational studies by the developers; no randomised trial was found.
– In FDA testing, about 88% of flags were false alarms (11.7% of flags were true sepsis).
– FDA test sensitivity (79.4%) missed the pre-set target of a lower bound above 80%.
– Not validated in children, maternity, recent post-surgery or unstabilised trauma patients.
– No CE marking or European use found; all published data come from US hospitals.
This profile is not medical advice or an endorsement. Hospitals should perform their own due diligence (regulatory compliance, GDPR, cybersecurity, integration, local clinical validation). How we evaluate
Every profile documents the precise indication, regulatory status, vendor-supported and independent evidence, real-world deployment, limits and dated sources.
A independent systematic review or prospective multicentre study · B at least one independent peer-reviewed study · C peer-reviewed, vendor-supported · D vendor claims only.
Vendors, authors and clinicians can request a correction. Corrections are dated on the profile.