RapidAI

By iSchemaView, Inc.
· United States
Modular stroke imaging software that triages suspected large vessel occlusion and intracranial hemorrhage on brain CT, with automated ASPECTS and perfusion volumes, as an aid to acute stroke teams.
Diagnostic
Neurology/Neurovascular
Production
FDA Cleared
Regulatory status
FDA Cleared
Deployment
Production
Clinical role
Adjunct (decision support)
Headquarters
United States
AI type
Diagnostic

Precise indication

RapidAI is a modular stroke imaging platform. Its modules fall into three roles.

Triage and notification (alerts the team, no diagnosis)

  • Rapid LVO: suspected large vessel occlusion (distal internal carotid artery and M1 segment of the middle cerebral artery) on CT angiography.
  • Rapid NCCT Stroke: intracranial hemorrhage and large vessel occlusion together, directly from non-contrast CT, without requiring CTA.
  • Rapid ICH: intracranial hemorrhage on non-contrast CT.

Diagnostic aid (read together with the radiologist)

  • Rapid ASPECTS: automated ASPECTS score from non-contrast CT, in adults with known MCA or ICA occlusion. It does not replace the radiologist’s interpretation.

Treatment selection (adults only)

  • Rapid Core (CTP/MR diffusion/mismatch): computes tissue volumes (infarct core and penumbra) to help select candidates for mechanical thrombectomy in the extended 6–24 hour window, based on the DAWN and DEFUSE 3 trials.

Regulatory status

FDA: all US clearances below are 510(k), Class II, applicant iSchemaView, Inc., Golden, CO.

Module Reference Date Status
United States — FDA 510(k)
Rapid ICH K193087
QAS, 21 CFR 892.2080
31 Mar 2020 To check (note 1)
Rapid ASPECTS K200760
POK, 21 CFR 892.2060
26 Jun 2020 Confirmed (FDA database)
Rapid LVO K200941
QAS, 21 CFR 892.2080
9 Jul 2020 Confirmed (FDA database)
Rapid LVO (later version) K221248 31 May 2022 Confirmed (FDA database)
Rapid NCCT Stroke K222884
QAS, 21 CFR 892.2080
2 Mar 2023 Confirmed (FDA database)
AngioFlow (perfusion in the angiography suite) 510(k), number not checked Announced May 2024 Vendor-claimed
Rapid DeltaFuse, Rapid LMVO, Rapid MLS, Rapid OH, Rapid Aortic Five clearances, numbers not checked Announced 25 Nov 2025 Vendor-claimed
European Union / Europe
Rapid ASPECTS Described as CE-marked at European launch (ESOC)
Likely MDD (interpretation, to check; note 2)
May 2018 Vendor-claimed
AngioFlow Europe (unspecified): vendor CEO, “already registered and used in Japan and Europe” – Vendor-claimed (note 3)
United Kingdom
RapidAI NICE HTG708 2024 Conditional NHS use (note 4)

Notes

  • Note 1 — Rapid ICH, K193087: from a third-party 510(k) database; official FDA page not retrieved, to check.
  • Note 2 — Rapid ASPECTS, EU, 2018: the reading as former Directive 93/42/EEC (MDD), not MDR, is inferred from the date only (interpretation, to check).
  • Note 3 — AngioFlow, Europe: the source does not mention CE marking, class or regulation.
  • Note 4 — RapidAI, UK: NICE HTG708 lists RapidAI among products that can be used in the NHS only while more evidence is generated, with DTAC approval and healthcare professional review; CE/UKCA class not stated in the NICE pages read.
  • No MDR (EU) 2017/745 certificate, notified body or EUDAMED registration was found for any Rapid module in public sources.
  • Current CE status, class and notified body must be obtained from the manufacturer (certificate copy) before any CE claim is stated.

Clinical evidence

Vendor-supported

Regulatory submission study (Rapid ASPECTS, K200760): concurrent-read crossover study, 50 cases, 8 readers of mixed specialties; software assistance added 0.425 regions of agreement per scan (95% CI 0.11-0.74); non-neuroradiologist agreement rose from 73.6% to 79.8% (p<0.0001). Summary from a third-party 510(k) database, not re-read on the FDA PDF. Vendor-claimed performance (Rapid LVO, clearance press release, 13 August 2020): 97% sensitivity, 96% specificity; press-release figures, not peer-reviewed. "DUEL" study: Sachdev H, Hudson A, Ong K, Marklein S, Flores M, AJNR, published online 14 May 2026, DOI 10.3174/ajnr.A9418. Single comprehensive stroke centre (Good Samaritan Hospital, HCA, San Jose, USA), 1,589 consecutive stroke alerts, 1,523 analysed, 147 confirmed LVOs. Rapid LVO 98% sensitivity (144/147) and about 94% specificity versus Viz.ai 73.5% (108/147) and about 91% (p<0.0001 and p=0.004). Promoted by the vendor; funding and conflict-of-interest statements not verified on the full article.

Independent

Lambert J, Demeestere J, et al. (University Hospitals Leuven / KU Leuven, Belgium), AJNR 2023;44(8):894-900, DOI 10.3174/ajnr.A7956: retrospective single-centre study, 174 thrombectomy-treated anterior-circulation stroke patients (175 non-contrast CTs), 5 raters. RAPID ASPECTS alone: weighted kappa 0.59 versus the criterion standard, similar to human raters (0.38-0.76). With software assistance, accuracy improved (AUC 0.72 to 0.82, p=0.004) and inter-rater agreement improved (Fleiss kappa 0.15 to 0.23 overall; 0.39 to 0.55 for dichotomized scores); for the dichotomized thrombectomy threshold, most individual rater improvements were not significant. Author-stated limits: thrombectomy-only population (low ASPECTS under-represented), criterion-standard readers had access to the software output, one vendor tested, heterogeneous referral CT protocols.
CDA-AMC (Canada’s Drug Agency) HTA, expert panel meeting 12 September 2024, recommendation December 2024: 2 cohort and 11 diagnostic accuracy studies, only 3 testing RapidAI as intended (alongside clinicians); standalone LVO sensitivity about 62-96% and specificity about 65-98% across 10 studies; ICH 92% sensitivity (95% CI 78-98) and 100% specificity (95% CI 98-100) from one low-certainty study; no evidence on mortality, length of stay, quality of life, patient harms or cost-effectiveness; all studies in centres with stroke expertise. Where implemented: use only alongside clinician interpretation and generate evidence; for new implementation: no recommendation for or against.
NICE HTG708 (UK, 2024): observational time-to-treatment studies including RapidAI judged at high risk of bias; use allowed only with healthcare professional review while evidence is generated.

Real-world deployment

Vendor claim (2026): used in 2,500+ hospitals globally. Earlier vendor figure (March 2021): over 1,800 hospitals in over 60 countries. Both are vendor-reported and not independently audited. Named sites confirmed by third parties are NHS trusts in the West Midlands, UK (Queen Elizabeth Hospital Birmingham as hub, Hereford County Hospital, Sandwell and City Hospitals, Worcestershire Royal Hospital, 2023-2024). No named routine deployment in Belgium or another EU country was found.

Interoperability

Vendor-claimed: standard integration via DICOM, HL7 and FHIR plus custom APIs; DICOM-based worklist prioritization or integration with worklist providers; AI outputs surfaced alongside native images in PACS; integration with RIS and EHR systems; described as vendor-agnostic and built for hybrid environments (“Edge Cloud” option). No DICOM conformance statement, HL7 version or FHIR resource list was found in public sources; confirm these, the supported PACS/RIS/EHR vendors and the cloud/on-premise architecture with the vendor before deployment.

European status

Belgium: no routine clinical deployment found in public sources; University Hospitals Leuven used RAPID ASPECTS in a retrospective research study (Lambert et al., AJNR 2023). United Kingdom (third-party, NHS trust publications): Queen Elizabeth Hospital Birmingham (University Hospitals Birmingham, West Midlands Stroke Network hub, June 2023); Hereford County Hospital (Wye Valley NHS Trust, December 2023); Sandwell and City Hospitals (Sandwell and West Birmingham NHS Trust, January 2024); Worcestershire Royal Hospital (Worcestershire Acute Hospitals NHS Trust, 2024). No named hospital in the EU-27 found.

Editorial verdict

Limits and uncertainties

– Evidence certainty rated low to very low by an independent HTA body (CDA-AMC, 2024); NICE (2024) allows NHS use only while more evidence is generated.
– Standalone LVO detection varies widely across studies (sensitivity about 62-96%, specificity about 65-98%); most accuracy studies tested the software standalone rather than as intended alongside clinicians.
– No independent evidence on mortality, length of stay, quality of life, patient harms or cost-effectiveness.
– Evidence comes from centres with stroke expertise; performance in smaller or less-resourced hospitals is untested.
– No public data on training-data representativeness or subgroup performance (age, sex, ethnicity); risk of automation bias noted by CDA-AMC.
– Triage modules (LVO, ICH, NCCT Stroke) are notification tools, not diagnostic devices: absence of an alert does not rule out the finding.
– Rapid ASPECTS is a concurrent-read aid only; independent agreement with expert consensus is moderate (weighted kappa 0.59).
– Current EU MDR certificate, risk class and notified body not publicly confirmed; EU data hosting not documented.
– No named routine deployment in Belgium or another EU-27 country found; European evidence of use is mainly from NHS England.

Sources

https://www.rapidai.com/ ; https://www.mobihealthnews.com/news/rapidai-receives-fda-510k-identify-potential-pulmonary-embolism https://radiologybusiness.com/topics/healthcare-management/healthcare-policy/rapidai-groundbreaking-fda-neuroimaging-stroke https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/POK/K200760 https://www.businesswire.com/news/home/20200813005190/en/RapidAI-Receives-FDA-Clearance-Rapid-LVO-Identification https://www.businesswire.com/news/home/20230418005344/en/RapidAI-Receives-First-and-Only-FDA-510k-Clearance-of-Non-Contrast-CT-Imaging-Product-to-Accelerate-Acute-Stroke-Triage https://www.regdatalab.com/510k/K222884 https://www.businesswire.com/news/home/20251125845646/en/RapidAI-Extends-the-Reach-of-Deep-Clinical-AI-with-Five-New-FDA-Clearances https://cda-amc.ca/sites/default/files/hta-he/OP0556_FINAL-Recommendation.pdf https://www.rapidai.com/blog/duel-study-published-in-ajnr-real-world-data-on-rapidai-lvo-detection https://itnonline.com/node/87117 https://www.rapidai.com/about-us Regulator — FDA, 510(k) K200941 Rapid LVO, accessdata.fda.gov, July 2020 https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K200941 Regulator — FDA, 510(k) K221248 Rapid LVO, accessdata.fda.gov, May 2022 https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K221248 Regulator — FDA, 510(k) K200760 Rapid ASPECTS, accessdata.fda.gov, June 2020 https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K200760 Regulator — FDA, 510(k) K222884 Rapid NCCT Stroke, accessdata.fda.gov, March 2023 https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K222884 Database — 510kdatabase.net, K193087 Rapid ICH, March 2020 https://www.510kdatabase.net/k193087/ Press release (vendor) — RapidAI, FDA clearance for perfusion imaging in the angiography suite (AngioFlow), rapidai.com, May 2024 https://www.rapidai.com/press-release/rapidai-secures-fda-clearance-for-perfusion-imaging-in-the-angiography-suite Press release (vendor) — iSchemaView, iSchemaView Introduces Rapid ASPECTS (CE mark, ESOC), rapidai.com, May 2018 https://www.rapidai.com/press-release/ischemaview-introduces-rapid-aspects News — Endovascular Today, iSchemaView’s Rapid ASPECTS Launched in Europe, May 2018 https://evtoday.com/news/ischemaviews-rapid-aspects-launched-in-europe Vendor — RapidAI, Workflow integration, rapidai.com, 2026 https://www.rapidai.com/platform/workflow-integration Study — Sachdev H et al., DUEL study (Rapid LVO vs Viz LVO), AJNR, May 2026 https://www.ajnr.org/content/early/2026/05/14/ajnr.A9418.long News — ITN, RapidAI Expands AI-Powered Vascular Imaging Globally (vendor figures), itnonline.com, March 2021 https://www.itnonline.com/content/rapidai-expands-ai-powered-vascular-imaging-globally Study — Lambert J, Demeestere J et al., Performance of Automated ASPECTS Software and Value as a CAD Tool, AJNR (PMC), August 2023 https://pmc.ncbi.nlm.nih.gov/articles/PMC10411841 Study — Lambert J et al., Performance of Automated ASPECTS Software, AJNR 44(8):894-900, August 2023 https://www.ajnr.org/content/44/8/894 HTA — CDA-AMC, RapidAI for Stroke Detection, Expert Panel recommendation, Canadian Journal of Health Technologies, December 2024 https://www.canjhealthtechnol.ca/index.php/cjht/article/download/OP0556-01/2277?inline=1 HTA — NICE, HTG708 AI-derived software to help clinical decision making in stroke (UK Government Web Archive copy), January 2024 updated May 2024 https://tnaqa.mirrorweb.com/ukgwa/20260505160951mp_/https://www.nice.org.uk/guidance/htg708 HTA — NICE, HTG708 Recommendations (UK Government Web Archive copy), May 2024 https://tnaqa.mirrorweb.com/ukgwa/20260505160951mp_/https://www.nice.org.uk/guidance/htg708/chapter/1-Recommendations HTA — NICE, HTG708 Committee discussion (UK Government Web Archive copy), May 2024 https://tnaqa.mirrorweb.com/ukgwa/20260505160951mp_/https://www.nice.org.uk/guidance/htg708/chapter/committee-discussion Hospital — University Hospitals Birmingham NHS FT, AI software to help clinicians save the lives of stroke patients in Birmingham, June 2023 https://www.uhb.nhs.uk/news-and-events/news/artificial-intelligence-software-to-help-clinicians-save-the-lives-of-stroke-patients-in-birmingham/ Hospital — Wye Valley NHS Trust, AI used for the first time at Hereford County Hospital for stroke patients, December 2023 https://www.wyevalley.nhs.uk/news-events/trust-newsroom/2023/december/ai-used-for-the-first-time-at-hereford-county-hospital-to-determine-care-for-stroke-patients.aspx Hospital — Sandwell and West Birmingham NHS Trust, Artificial intelligence aids faster stroke diagnosis, January 2024
Artificial intelligence aids faster stroke diagnosis
Hospital — Worcestershire Acute Hospitals NHS Trust, AI software helping stroke patients, 2024 https://www.worcsacute.nhs.uk/artificial-intelligence-software-helping-stroke-patients-at-worcestershire-acute-hospitals-nhs-trust/ Regulation — EU, Regulation (EU) 2024/1689 (AI Act), EUR-Lex, July 2024 https://eur-lex.europa.eu/eli/reg/2024/1689/oj

This profile is not medical advice or an endorsement. Hospitals should perform their own due diligence (regulatory compliance, GDPR, cybersecurity, integration, local clinical validation). How we evaluate

At a glance
Vendor
iSchemaView, Inc.
Headquarters
United States
AI type
Diagnostic
Clinical role
Adjunct (decision support)
CE marking
Legacy CE (MDD) claimed for Rapid ASPECTS, 2018; MDR status not found
Notified body
EU AI Act
High-risk (Annex I via MDR), if MDR-certified; to confirm
Data hosting
Last verified
Reviewed by
How we evaluate

Every profile documents the precise indication, regulatory status, vendor-supported and independent evidence, real-world deployment, limits and dated sources.

A independent systematic review or prospective multicentre study · B at least one independent peer-reviewed study · C peer-reviewed, vendor-supported · D vendor claims only.

Read the methodology →

Spotted an error?

Vendors, authors and clinicians can request a correction. Corrections are dated on the profile.