Use case · Capital Region, Denmark

AI as a first reader in breast cancer screening

The Capital Region of Denmark's public screening programme uses a commercial AI system to sort mammograms. Exams the AI rates as likely normal are read by one radiologist instead of two; all others keep double reading, with AI support.

Commercial productRadiologyBreast cancer screeningTranspara (ScreenPoint Medical)PACSIn production since 2021
How this case was documented: from published sources, mainly the programme's peer-reviewed evaluation (Radiology, 2024) and a 2026 conference presentation by the programme. It has not yet been reviewed by the Danish team; we will update it with their input.
33.5%fewer radiologist reads in the first year with AI
0.82%cancer detection rate, up from 0.70% before AI
1.63%false-positive rate, down from 2.39% before AI
58,246women screened in the first year with AI

01The problem

Denmark invites women aged 50 to 69 to a mammogram every two years. In the Capital Region alone, about 75,000 women accept each year, and every exam is read independently by two breast radiologists.

  • Too few readers. The region said publicly that it lacks breast radiologists to sustain double reading.
  • Most exams are normal. Radiologists spend most of their reading time on mammograms that show no cancer.
  • False alarms are costly. Before AI, 3.09% of women were recalled, and most of them did not have cancer. The programme estimates that one diagnostic work-up costs as much radiologist time as 30 to 50 screening reads.

02The solution

The region bought Transpara, a commercial AI system from the Dutch company ScreenPoint Medical, supplied through the Danish company Human Bytes. It has been in routine use since 18 November 2021.

  • Tested locally first. Before buying, the programme ran retrospective simulations of the AI on 114,421 of its own past screening exams (2014–2015), with two-year follow-up.
  • What it does: it scores each mammogram for the likelihood of cancer and marks suspicious areas. It was not trained on the region's own images, according to the programme.
  • How the score is used: exams with a low score (5 or less on a 1–10 scale, raised to 7 or less from 3 May 2022) are single-read by one of 19 senior full-time breast radiologists. All other exams are double-read, with the AI marks as decision support. About two-thirds of exams fall in the single-read group.
  • Role: the AI never recalls a woman on its own. A radiologist makes every final decision, and disagreements go to consensus or a third reader.
  • A known gap: the AI does not look at the woman's previous mammograms; radiologists do.

03Integration & governance

The AI was added to the existing reading workflow, not built around it: radiologists see its score in the PACS and keep full control of recall decisions.

AreaWhat the Capital Region did
Workflow integrationThe AI score is shown in the PACS at the end of each exam. Integration with the RIS is not described in public sources.
ProcurementAbout three months from procurement to go-live, run jointly by the regional IT centre, the supplier and the screening clinicians.
Validation before go-liveTwo retrospective simulation studies on the programme's own past exams, run with the University of Copenhagen.
Safety rulesDouble reading kept for every exam not rated low-risk. No recall by AI alone. A radiologist signs off every exam.
Threshold changesThe single-read threshold was raised once, on 3 May 2022. The reason has not been published.
Medical deviceTranspara is CE-marked and FDA-cleared, according to the programme. Class and notified body are not stated in public sources.
Data protectionNot described in public sources (hosting, data processing agreement, DPIA).
MonitoringScreening quality is tracked through the Danish quality database for mammography screening: recall, false positives, detection rate and, later, interval cancers.
Clinician involvementThe project was led by the screening programme itself, with its breast radiologists and academic partners.

04Measured results

In the first year with AI, radiologists read a third fewer exams, more cancers were found and fewer women were recalled for nothing. These are early indicators from a before-and-after comparison, not a randomised trial.

Study design: 60,751 women screened without AI (October 2020 – November 2021) compared with 58,246 women screened with AI (November 2021 – October 2022), followed for at least 180 days.

IndicatorBefore AIWith AI
Radiologist readsAll exams double-read33.5% fewer reads (38,977 of 116,492 saved)
Recall rate3.09%2.46%
Cancer detection rate0.70%0.82%
False-positive rate2.39%1.63%
Positive predictive value of a recall22.6%33.6%
Small cancers (1 cm or less)36.6%44.9%
Node-negative cancers76.7%77.8% (no significant change)

A caution on the comparison: the time between two screenings was longer in the AI period (median 993 days against 845). A longer interval can by itself raise the number of cancers found, so part of the gain may not come from the AI.

Low-risk group: up to the end of 2022, women in the single-read group had a recall rate of 0.40%. Of the 215 recalled, 14 had cancer, according to the programme.

Not measured yet: interval cancers (cancers found between two screenings), overdiagnosis, effect on mortality, and cost.

05Lessons learned

The programme used AI to move radiologists' time from normal exams to the uncertain ones, without letting the AI decide alone, and screening quality held up in the first year.

What worked, according to the programme

  • Testing the AI on the programme's own past exams before buying it.
  • Keeping a radiologist on every exam and double reading for every exam not rated low-risk.
  • A short, joint procurement with the regional IT centre, the supplier and the clinicians.
  • Tracking recall closely: a small rise in recalls quickly cancels the time saved.

What did not, or not yet

  • Interval cancers, the key safety measure, are still being collected and analysed.
  • The AI cannot compare with previous mammograms. All 14 cancers found in recalled low-risk women were new or changed since the last exam.
  • One region, one product and a before-and-after design limit what can be concluded. Longer screening intervals in the AI period blur the comparison.

Cost and effort: not published (licence, integration work, staff time).